KWS-3 'VerySANS'

focusing VSANS diffractometer operated by JCNS@MLZ

User Tools

Site Tools

se:rheo

Rheo-SANS

Rheo-SANS :: dynamic SANS sample environments :: Rheometerss in conjunction with SANS/VSANS


JCNS-MLZ SANS group has two Anton Paar MCR501 and MCR702 series rheometers available to simultaneously probe the shear rheology and the material microstructure in conjunction with SANS/VSANS. The flow geometry consists of a Couette type cell. Quartz and titanium cylinders are available with flow gaps ranging from 0.5 mm to 1.5 mm. The sample volumes range from 3 ml for the smallest combination up to 14 ml for the largest. A range of rheological tests can be performed in conjunction with SANS/VSANS, from basic steady shear mode to more advanced time-dependent measurement such as large-amplitude oscillatory shear (LAOS) mode.

Technical specifications
Code: JRH-1-JCNS
Torque Range 0.5 nNm - 230 mNm
Torque Resolution 0.1 nNm
Shear rate: 0.001 to 5000 s-1
Angular Frequency Range 10-7 rad/s to 628 rad/s
Temperature Range −40 to 200°C
Quartz Cup Diameter(s) 50 mm
Titanium Cup Diameter(s) 50 mm, 30 mm, 29 mm, 28 mm
Quartz Bob Diameter(s) 49 mm, 29 mm
Titanium Cup Diameter(s) 49 mm, 48 mm, 29 mm, 28 mm, 27 mm
Sample Volume, 50mm cup-49mm 11 ml
Sample Volume, 50mm cup-48mm 14 ml
Sample Volume, 30mm cup-29mm 3.75 ml
Sample Volume, 29mm cup-28mm 3.5 ml
Sample Status liquid samples


se/rheo.txt · Last modified: 2023/06/15 12:04 by Vitaliy Pipich